Home|Journals|Articles by Year|Audio Abstracts
 

Original Research



A comparison study of routine coagulation screening tests (PT and APTT) by three automated coagulation analyzers

Nitinkumar N Rathod, Sukesh C Nair, Joy Mammen, Surinder Singh.




Abstract

Background: In recent times, growing coagulation test volume and constricted personnel budgets have enhanced interest in automated coagulation analyzers.

Objective: (i) To compare the reliability of routine coagulation test [prothrombin time (PT) and activated partial thromboplastin time (APTT)] using mechanical, photo-optical, and nephelemetric methods by three automated coagulation analyzers. (ii) To evaluate the performance of a newly installed fully automatic coagulation analyzer [STA Compact Max (Stago)] and compare the consistency of its testing results with the confirmed clinical automatic coagulation analyzer at our department (Sysmex and ACL Top).

Materials and Methods: Trisodium citrated (3.2%) 60 blood samples, which came to special (coagulation) laboratory with request of PT and APTT (with or without anticoagulant, abnormal, and normal/controls), Transfusion Medicine and Immunohematology Department, Christian Medical College, Vellore, Tamil Nadu, India, were included into study randomly. Sample were run on Sysmex CS2000i (Sysmex Corporation, Japan), photo-optical clot detection; ACL Top (Instrumentation Laboratory, USA), nephelometry clot detection method; and STA Compact Max (Stago, USA), viscosity-based (mechanical) clot detection.

Result: Correlation was determined using Bland and Altman analysis, which demonstrated a good agreement between Sysmex CS2000i, ACL Top, and STA Compact Max for PT and APTT. A total of 10 samples (16.7%) with visually observed interferences were identified and tested on all three analyzers and showed good agreement with optical and mechanical methods.

Conclusion: This study showed good agreement between newly installed STA Compact Max based on mechanical endpoint detection method with already standardized Sysmex CS 2000i and ACL Top, which work on photo-optical endpoint detection method, for evaluation of screening coagulation tests such as PT and APTT even in case of variables such as partially lysed/icteric samples. Three automated analyzers can be used interchangeably.

Key words: Coagulation, automated analyzer, coagulation screening tests






Full-text options


Share this Article


Online Article Submission
• ejmanager.com




ejPort - eJManager.com
Refer & Earn
JournalList
About BiblioMed
License Information
Terms & Conditions
Privacy Policy
Contact Us

The articles in Bibliomed are open access articles licensed under Creative Commons Attribution 4.0 International License (CC BY), which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.