ADVERTISEMENT

Home|Journals|Articles by Year|Audio Abstracts
 

Original Research



Comparison of the sodium, potassium, chloride and glucose levels measured by a blood gas analyzer and an autoanalyzer

Omer Kaya, Kamile Yucel.



Abstract
Download PDF Cited by 0 ArticlesPost

Aim: A blood gas analyzers (BGA) are vital equipment frequently used in emergency departments and intensive care units. It is clinically important that the measurements of a BGA and an autoanalyzer (AA) provide equivalent results, which is confirmed by their proximitiy to the absolute value. This study aimed to compare the sodium (Na+), potassium (K+), chloride (Cl-) and glucose values in venous blood samples measured with a BGA and a standard AA with external quality control values.
Material and Methods: The results of patients that presented to our emergency department between April 1, 2019 and July 1, 2019 and underwent the measurements of Na+ (n = 5,908), K+ (n = 5,755), Cl- (n = 5,101) and glucose (n = 5,871) simultaneously by BGA and AA were retrospectively compared.
Results: In the Spearman correlation analysis between the two measurements, the correlation coefficient (r) was found as 0.78, 0.88, 0.89 and 0.97 for Na+, K+, Cl- and glucose, respectively. According to the Bland-Altman analysis, in the comparison of Na+, K+, Cl- and glucose values, the average bias percentages at the 95% confidence interval were -0.8 (4.8 to -6.4), -9 (8.6 to -26.5), -0.2 (6.2 to -6.5), and 0.3 (21.8 to -21.3), respectively.
Conclusion: We concluded that the Na+ and Cl- results obtained from BGA can be used instead of those obtained from an AA; however, the blood gas K+ and glucose results of the former cannot replace those of the latter.

Key words: Acid-base balance; autoanalyzer; blood gas analysis; electrolytes; glucose







Bibliomed Article Statistics

52
41
33
42
27
21
29
31
33
25
15
8
R
E
A
D
S

6

12

6

6

9

9

9

16

14

16

6


D
O
W
N
L
O
A
D
S
070809101112010203040506
20252026

Full-text options


Share this Article


Online Article Submission
• ejmanager.com




ejPort - eJManager.com
Author Tools
About BiblioMed
License Information
Terms & Conditions
Privacy Policy
Contact Us

The articles in Bibliomed are open access articles licensed under Creative Commons Attribution 4.0 International License (CC BY), which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.