Introducing SRF in Bismuth-Silicon and Polyurethane shields for breast dose reduction during chest CT

Parinaz Mehnati, Vida Sargazi, Zainab Yazdi Sotoodeh, Afshin Nejadjahantigh, Masoome Farsizaban.




Abstract

Background: This study introduces Shield Reduction Factor (SRF) for breast dose decline using two new Bismuth-polymer composites shields in chest CT.
Method and Materials: SRF of Bismuth -Silicon (Bi-Si) and -Polyurethane (Bi-PU) were measured after chest CT by a multi-detector CT (MDCT, 16 detectors) and a single-detector CT (SDCT) for breast at spiral and axial modes using female chest phantom.
Result: Higher effectiveness was observed in the use of the Bi-Si shield with an SRF of 1.25 and 1.66, and in the use of the Bi-PU shield with an SRF of 2.31 and 2.06 in SDCT and MDCT (100 mAs), axial mode, respectively. Experiments by 80 mAs at spiral/axial mode induced lower SRF by Bi-Si and Bi-PU shields.
Conclusion: This study showed that Bi-Si and Bi-PU shields are new composites, light and economical, carrying promising potentials for breast dose reduction during chest CT in spiral and axial modes at 80 and 100 mAs, single and 16 MDCT.

Key words: Bi-Si ,Bi-PU shields ,breast dose , SDCT, MDCT






Full-text options


Share this Article


Online Article Submission
• ejmanager.com




ejPort - eJManager.com
Refer & Earn
JournalList
About BiblioMed
License Information
Terms & Conditions
Privacy Policy
Contact Us

The articles in Bibliomed are open access articles licensed under Creative Commons Attribution 4.0 International License (CC BY), which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.